More work on EK-VSTAA-MG-001 chapter 2
This commit is contained in:
Binary file not shown.
@@ -36,6 +36,7 @@ mode of operation each prompt represents.
|
|||||||
\begin{table}[H]
|
\begin{table}[H]
|
||||||
\caption{Prompts}
|
\caption{Prompts}
|
||||||
\label{table:2-1}
|
\label{table:2-1}
|
||||||
|
\small
|
||||||
\begin{tabularx}{\textwidth}{c l}
|
\begin{tabularx}{\textwidth}{c l}
|
||||||
\hline
|
\hline
|
||||||
\textbf{Prompt} & \textbf{Mode of operation}\\
|
\textbf{Prompt} & \textbf{Mode of operation}\\
|
||||||
@@ -203,27 +204,20 @@ installed, but its ROM is destroyed and the Option device must be replaced.
|
|||||||
Only TESTS 1 through 4 can have an asterisk after them. Figure 2-2 lists
|
Only TESTS 1 through 4 can have an asterisk after them. Figure 2-2 lists
|
||||||
the symbols that can appear in between the tests and what they indicate.
|
the symbols that can appear in between the tests and what they indicate.
|
||||||
|
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-1}{Example of Power-up Tests Screen Display}
|
||||||
\caption{Example of Power-up Tests Screen Display}
|
|
||||||
\begin{verbatim}
|
|
||||||
KA410-A V1.0
|
KA410-A V1.0
|
||||||
F...E...D...C...B...A...9...8...7...6...5...4_..3_..2_..1_..
|
F...E...D...C...B...A...9...8...7...6...5...4_..3_..2_..1_..
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
|
||||||
|
|
||||||
Figure 2-2 lists the definitions of the symbols that appear between the test
|
Figure 2-2 lists the definitions of the symbols that appear between the test
|
||||||
numbers in the power-up test countdown.
|
numbers in the power-up test countdown.
|
||||||
|
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-2}{Power-up Symbols Defined}
|
||||||
\caption{Power-up Symbols Defined}
|
|
||||||
\begin{verbatim}
|
|
||||||
... Device tested successfully or has a soft error
|
... Device tested successfully or has a soft error
|
||||||
?.. Device has a hard error
|
?.. Device has a hard error
|
||||||
_.. Device not installed or not tested
|
_.. Device not installed or not tested
|
||||||
*.. Device installed but its ROM is destroyed
|
*.. Device installed but its ROM is destroyed
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
|
||||||
\label{figure:2-1}
|
|
||||||
|
|
||||||
If any hard errors (errors that indicate the device must be replaced for
|
If any hard errors (errors that indicate the device must be replaced for
|
||||||
proper operation) are found during power-up testing, a question mark is
|
proper operation) are found during power-up testing, a question mark is
|
||||||
@@ -248,23 +242,19 @@ explanation of how to use the configuration table.
|
|||||||
|
|
||||||
Each error summary consists of one or two question marks, a test number,
|
Each error summary consists of one or two question marks, a test number,
|
||||||
the ID number of the failed device, and an eight-digit error code. For ex-
|
the ID number of the failed device, and an eight-digit error code. For ex-
|
||||||
ample, in Figure 2-3, the first line of the error summary shows a hard error
|
ample, in \hyperref[figure:2-3]{Figure 2-3}, the first line of the error summary shows a hard error
|
||||||
for TEST F, a device ID number of 00B0, and an error code of 0001.F002.
|
for TEST F, a device ID number of 00B0, and an error code of 0001.F002.
|
||||||
The second line shows a soft error for TEST E, a device ID of 0040, and an
|
The second line shows a soft error for TEST E, a device ID of 0040, and an
|
||||||
error code of 0000.0005. Section ( 2.3.1.1) describes the error codes.
|
error code of 0000.0005. \hyperlink{subsubsection.2.3.1.1}{Section 2.3.1.1} describes the error codes.
|
||||||
\newpage
|
\newpage
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-3}{Example of Power-up Tests Screen Display with Errors.}
|
||||||
\caption{Example of Power-up Tests Screen Display with Errors.}
|
|
||||||
\begin{verbatim}
|
|
||||||
KA410-A V1.0
|
KA410-A V1.0
|
||||||
|
|
||||||
F?..E...0...C...B...A...9...8...7...6...5...4_..3_..2_..1_..
|
F?..E...0...C...B...A...9...8...7...6...5...4_..3_..2_..1_..
|
||||||
|
|
||||||
?? F 00B0 0001.F002
|
?? F 00B0 0001.F002
|
||||||
? E 0040 0000.0006
|
? E 0040 0000.0006
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
|
||||||
\label{figure:2-2}
|
|
||||||
|
|
||||||
If there is a fatal error in the NVR during power-up testing, the system stops
|
If there is a fatal error in the NVR during power-up testing, the system stops
|
||||||
testing the other devices and displays ?14 TOY ERR on the screen. When
|
testing the other devices and displays ?14 TOY ERR on the screen. When
|
||||||
@@ -275,6 +265,7 @@ the failing module. \hyperref[table:2-2]{Table 2-2} lists the LEDs and which mod
|
|||||||
\begin{table}
|
\begin{table}
|
||||||
\label{table:2-2}
|
\label{table:2-2}
|
||||||
\caption{Keyboard LEDs Deflned}
|
\caption{Keyboard LEDs Deflned}
|
||||||
|
\small
|
||||||
\begin{tabularx}{\textwidth}{l l}
|
\begin{tabularx}{\textwidth}{l l}
|
||||||
\hline
|
\hline
|
||||||
\textbf{Keyboard LED} & \textbf{Failing module}\\
|
\textbf{Keyboard LED} & \textbf{Failing module}\\
|
||||||
@@ -315,15 +306,14 @@ of the self-test and power-up tests and is updated each time self-test is run.
|
|||||||
The error codes for each device in the configuration table are explained in
|
The error codes for each device in the configuration table are explained in
|
||||||
the troubleshooting section for that individual device. Remember that the
|
the troubleshooting section for that individual device. Remember that the
|
||||||
configuration table contains the results of the self-test and power-up tests
|
configuration table contains the results of the self-test and power-up tests
|
||||||
and not the results of the system exerciser. Figure 2-4 shows an example
|
and not the results of the system exerciser. \hyperref[figure:2-4]{Figure 2-4} shows an example
|
||||||
of the configuration table and for an explanation of the configuration table,
|
of the configuration table and for an explanation of the configuration table,
|
||||||
see Paragraph 2.5.1.
|
see \hyperlink{subsection.2.5.1}{Paragraph 2.5.1}.
|
||||||
|
|
||||||
|
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-4}{Example of the Configuration Table}
|
||||||
\caption{Example of the Configuration Table}
|
|
||||||
\begin{verbatim}
|
|
||||||
>>> TEST 50
|
>>> TEST 50
|
||||||
|
|
||||||
KA410-A V1.0
|
KA410-A V1.0
|
||||||
ID 08-00-2B-02-CF-A4
|
ID 08-00-2B-02-CF-A4
|
||||||
|
|
||||||
@@ -344,10 +334,8 @@ ID 08-00-2B-02-CF-A4
|
|||||||
SYS 0000.0001
|
SYS 0000.0001
|
||||||
NI 0000.0001 V1.0
|
NI 0000.0001 V1.0
|
||||||
>>>
|
>>>
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
\newpage
|
||||||
\label{figure:2-3}
|
|
||||||
|
|
||||||
The most common good error code is 0000.0001. There are, however, some
|
The most common good error code is 0000.0001. There are, however, some
|
||||||
devices that use the first four digits in the error code to indicate the status
|
devices that use the first four digits in the error code to indicate the status
|
||||||
of the device and the last four digits to indicate the error found on the
|
of the device and the last four digits to indicate the error found on the
|
||||||
@@ -379,15 +367,13 @@ tested. \hyperref[table:2-3]{Table 2-3} lists the test numbers and the devices t
|
|||||||
numbers. Figure 2-5 shows an example of running self-test successfully on
|
numbers. Figure 2-5 shows an example of running self-test successfully on
|
||||||
the disk controller.
|
the disk controller.
|
||||||
|
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-5}{Example of Running Self-test on the Disk Controller}
|
||||||
\caption{Example of Running Self-test on the Disk Controller}
|
|
||||||
\begin{verbatim}
|
|
||||||
>>> TEST 7
|
>>> TEST 7
|
||||||
7...
|
7...
|
||||||
>>>
|
>>>
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
|
||||||
\label{figure:2-4}
|
\newpage
|
||||||
|
|
||||||
To test a group of devices, enter TEST followed by the test number of the
|
To test a group of devices, enter TEST followed by the test number of the
|
||||||
first device to be tested and then the test number of the last device to be
|
first device to be tested and then the test number of the last device to be
|
||||||
@@ -396,15 +382,11 @@ tested. Figure 2-6 shows an example of testing a group of devices. In Figure
|
|||||||
pick and choose which devices to test between C and 4, all tests between
|
pick and choose which devices to test between C and 4, all tests between
|
||||||
C and 4 are tested when entered as a group.
|
C and 4 are tested when entered as a group.
|
||||||
|
|
||||||
\begin{figure}[H]
|
\begin{ttfig}{2-6}{Example of Running a Series of Self-tests}
|
||||||
\caption{Example of Running a Series of Self-tests}
|
|
||||||
\begin{verbatim}
|
|
||||||
>>> TEST C 4
|
>>> TEST C 4
|
||||||
C...B...A...9...8...7...6...5...4_..
|
C...B...A...9...8...7...6...5...4_..
|
||||||
>>>
|
>>>
|
||||||
\end{verbatim}
|
\end{ttfig}
|
||||||
\end{figure}
|
|
||||||
\label{figure:2-5}
|
|
||||||
|
|
||||||
To test all devices, enter TEST F 1. The MicroVAX 2000 skips over the
|
To test all devices, enter TEST F 1. The MicroVAX 2000 skips over the
|
||||||
MONO video test (TEST F) since it does not use the video circuits.
|
MONO video test (TEST F) since it does not use the video circuits.
|
||||||
@@ -412,25 +394,26 @@ MONO video test (TEST F) since it does not use the video circuits.
|
|||||||
\begin{table}[H]
|
\begin{table}[H]
|
||||||
\caption{Self-test Commands}
|
\caption{Self-test Commands}
|
||||||
\label{table:2-3}
|
\label{table:2-3}
|
||||||
|
\small
|
||||||
\begin{tabularx}{\textwidth}{p{2cm} p{8cm}}
|
\begin{tabularx}{\textwidth}{p{2cm} p{8cm}}
|
||||||
\hline
|
\hline
|
||||||
\raggedright\textbf{Test\newline Number} & \textbf{Device\newline Tested} \\
|
\raggedright\textbf{Test\newline Number} & \textbf{Device\newline Tested} \\
|
||||||
\hline
|
\hline
|
||||||
1 & Option module (Network Interconnect module) (NI) \\
|
1 & Option module (Network Interconnect module) (NI) \\[0.5em]
|
||||||
2 & Option module (not available) \\
|
2 & Option module (not available) \\[0.5em]
|
||||||
3 & Option module (not available) \\
|
3 & Option module (not available) \\[0.5em]
|
||||||
4 & Option module (not available) \\
|
4 & Option module (not available) \\[0.5em]
|
||||||
5 & Interrupt Controller and ThinWire Ethernet ID ROM (SYS) \\
|
5 & Interrupt Controller and ThinWire Ethernet ID ROM (SYS) \\[0.5em]
|
||||||
6 & Tape Controller. (TPC) \\
|
6 & Tape Controller. (TPC) \\[0.5em]
|
||||||
7 & Disk conroller. (HDC) \\
|
7 & Disk conroller. (HDC) \\[0.5em]
|
||||||
8 & Interval timer. (IT) \\
|
8 & Interval timer. (IT) \\[0.5em]
|
||||||
9 & Floating point unit. (FP) \\
|
9 & Floating point unit. (FP) \\[0.5em]
|
||||||
A & Memory management unit. (MM) \\
|
A & Memory management unit. (MM) \\[0.5em]
|
||||||
B & Memory. (MEM) \\
|
B & Memory. (MEM) \\[0.5em]
|
||||||
C & DZ Controller. (DZ) \\
|
C & DZ Controller. (DZ) \\[0.5em]
|
||||||
D & Non-volatile RAM. (NVR) \\
|
D & Non-volatile RAM. (NVR) \\[0.5em]
|
||||||
E & Time-of-year dock. (CLK) \\
|
E & Time-of-year dock. (CLK) \\[0.5em]
|
||||||
F & Base video (MONO) (VAXstation 2000 only) \\
|
F & Base video (MONO) (VAXstation 2000 only) \\[0.5em]
|
||||||
\hline
|
\hline
|
||||||
\end{tabularx}
|
\end{tabularx}
|
||||||
\end{table}
|
\end{table}
|
||||||
@@ -472,17 +455,227 @@ that the configuration table contains the results of the self-test and power-up
|
|||||||
tests and not the results of the system exerciser. See Paragraph 2 5 1 for an
|
tests and not the results of the system exerciser. See Paragraph 2 5 1 for an
|
||||||
explanation of the configuration table.
|
explanation of the configuration table.
|
||||||
|
|
||||||
\begin{figure}[H]
|
\newpage
|
||||||
\caption{Example of a Self-test Error on the Disk Controller}
|
\begin{ttfig}{2-7}{Example of a Self-test Error on the Disk Controller}
|
||||||
\begin{verbatim}
|
|
||||||
>>> TEST 7
|
>>> TEST 7
|
||||||
7?..
|
7?..
|
||||||
84 FAIL
|
84 FAIL
|
||||||
>>>
|
>>>
|
||||||
\end{verbatim}
|
|
||||||
\end{figure}
|
|
||||||
\label{figure:2-6}
|
|
||||||
|
|
||||||
(You must display the configuration table to see the error code)
|
(You must display the configuration table to see the error code)
|
||||||
|
\end{ttfig}
|
||||||
|
|
||||||
|
\newpage
|
||||||
|
|
||||||
|
\subsection{System Exerciser Diagnostics}
|
||||||
|
|
||||||
|
The system exerciser simulates a worst-case situation test for each device
|
||||||
|
and checks how the device operates under these conditions. It does not use
|
||||||
|
the configuration table to list its results of the tests. The system exerciser
|
||||||
|
has a separate display that appears on the screen as the exerciser is running.
|
||||||
|
Any errors found are displayed in the exerciser display. When examining
|
||||||
|
the exerciser display, a single question mark in the far left column indicates
|
||||||
|
a soft error, a double question mark indicates a hard error, and the absence
|
||||||
|
of question marks indicate success. \hyperref[figure:2-8]{Figure 2-8} shows an example of the
|
||||||
|
exerciser display.
|
||||||
|
|
||||||
|
The system exerciser exercises most of the devices. However, some devices
|
||||||
|
such as the memory management unit (MM) and the interval timer (IT) are
|
||||||
|
tested through the testing of other devices and are not displayed. Of the
|
||||||
|
devices the exerciser does exercise, it runs each one sequentially until all
|
||||||
|
have been run once, then it runs them all at the same time (worst-case).
|
||||||
|
This type of testing usually finds any intermittent failures.
|
||||||
|
|
||||||
|
The system exerciser has two modes: customer mode and field service
|
||||||
|
mode. Customer mode system exerciser (TEST 0) does not use loopback
|
||||||
|
connectors and does not fully test all of the devices. Field service mode sys-
|
||||||
|
tem exerciser requires loopback connectors installed and removable media
|
||||||
|
from the maintenance kit inserted and loaded. You must initialize the floppy
|
||||||
|
diskette in the maintenance kit with a special diagnostic key so the exerciser
|
||||||
|
can perform write tests on the RX33. Refer to Paragraph 2.5.11 for more
|
||||||
|
Information on creating the special diagnostic keys. Once the floppy has
|
||||||
|
been initialized, load it into the RX33 (if a full read/write test of the RX33 is
|
||||||
|
necessary) before you run the system exerciser in field service mode. You
|
||||||
|
must also initialize the COMPACTape cartridge in the maintenance kit with
|
||||||
|
a special diagnostic key so the exerciser can perform read/write tests on
|
||||||
|
the TK50. Refer to Paragraph 2.5.11 for more information on creating the
|
||||||
|
special diagnostic keys. Once the COMPACTape cartridge has been initial
|
||||||
|
ized, load it into the TK50 (if a full read/write test of the TK50 is necessary)
|
||||||
|
before you run the system exerciser in field service mode. If the RX33 or
|
||||||
|
the TK50 are not loaded with the special-key media, the system exerciser
|
||||||
|
does not do destructive writes to them and tests them the same as it does
|
||||||
|
during the customer mode system exerciser. This special-key on the media
|
||||||
|
prevents the exerciser from accidentally destroying data on the customers
|
||||||
|
floppy diskette or COMPACTape cartridge. The field service mode system
|
||||||
|
exerciser is available in a run once (TEST 101) and a run forever (TEST 102)
|
||||||
|
configuration.
|
||||||
|
|
||||||
|
\newpage
|
||||||
|
|
||||||
|
\subsubsection{System Exerciser Diagnostic Commands}
|
||||||
|
\hyperref[table:2-4]{Table 2-4} lists the System exerciser diagnostic commands.
|
||||||
|
|
||||||
|
\begin{table}[H]
|
||||||
|
\label{table:2-4}
|
||||||
|
\caption{System Exerciser Diagnostic Commands}
|
||||||
|
\small
|
||||||
|
\begin{tabularx}{\textwidth}{p{0.2\textwidth} p{0.8\textwidth}}
|
||||||
|
\hline
|
||||||
|
\textbf{Test Commands} & \textbf{Description of Commands} \\
|
||||||
|
\hline
|
||||||
|
|
||||||
|
0 & Runs customer mode System exerciser. It exercises
|
||||||
|
each device once sequentially, then exercises them
|
||||||
|
simultaneously, and stops when the slowest device
|
||||||
|
finishes its second pass. No loopback connectors
|
||||||
|
and no removable media required. \\
|
||||||
|
|
||||||
|
101 & Runs field service system exerciser. It exercises
|
||||||
|
each device once sequentially, then exercises them
|
||||||
|
simultaneously, and stops when the slowest device
|
||||||
|
finishes its second pass. Do not stop the exerciser
|
||||||
|
before it finishes exercising every device twice
|
||||||
|
(second pass). Loopbacks and removable media required. \\
|
||||||
|
|
||||||
|
102 & Runs field Service System exerciser. It exercises each
|
||||||
|
device once sequentially and then exexcises them
|
||||||
|
simultaneously until you enter a \keystroke{CTRL}-\keystroke{C}.
|
||||||
|
Note that the exerciser takes up to thirty seconds
|
||||||
|
to stop after you enter \keystroke{CTRL}-\keystroke{C}. Do
|
||||||
|
not stop the exerciser until every device is exercised
|
||||||
|
twice (second pass). Also, do not press the halt button
|
||||||
|
to stop the exerciser. Loopbacks and removable media required. \\
|
||||||
|
\hline
|
||||||
|
\end{tabularx}
|
||||||
|
\end{table}
|
||||||
|
|
||||||
|
When the exerciser is started, PRA0 is displayed and the monitor connected
|
||||||
|
to the video port on the VAXstation 2000 blinks white and black several
|
||||||
|
times while the monochrome circuits are being tested. The results of the
|
||||||
|
MONO tests are then displayed on the console screen. MicroVAX 2000
|
||||||
|
does not use the MONO circuits and, therefore, does not test them. The
|
||||||
|
DZ test results are the first to be displayed on the console screen. The rest
|
||||||
|
of the devices are then exercised one at a Urne and the results are listed
|
||||||
|
on the display. The console displays the results of each device until the
|
||||||
|
last device is finished testing. When the last device is done, the exerciser
|
||||||
|
starts running all devices together at the same time. When this happens, the
|
||||||
|
monitor connected to the video port starts blinking again until the slowest
|
||||||
|
device finishes testing. On the MicroVAX 2000, the console terminal holds
|
||||||
|
the first pass display until the slowest device is done testing then displays a
|
||||||
|
new exerciser display. The console screen (VAXstation 2000 and MicroVAX
|
||||||
|
2000) displays the results of each device tested. This display stays on the
|
||||||
|
screen for about 10 seconds, then the exerciser starts running all devices
|
||||||
|
together again if TEST 102 was entered. If TEST 0 or TEST 101 was entered,
|
||||||
|
the exerciser stops after the slowest device is done testing. The halt message
|
||||||
|
is displayed when the exerciser is stopped.
|
||||||
|
|
||||||
|
To run the field service mode system exerciser on VAXstation 2000, install a
|
||||||
|
loopback (p/n 29-24795) on the 25-pin communication port and a loopback
|
||||||
|
(p/n 29-24794) on the 9-pin printer port. Insert and load the floppy diskette
|
||||||
|
and TK50 COMPACTape cartridge from the maintenance kit if the system
|
||||||
|
has these devices installed. Run the exerciser by entering TEST 101 or TEST
|
||||||
|
102. If you are using the diagnostic console device, you will get errors on
|
||||||
|
the DZ line since there is no loopback for the video port.
|
||||||
|
|
||||||
|
To run the field service mode system exerciser on MicroVAX 2000, install
|
||||||
|
a loopback (p/n 29-24795) on the 25-pin communication port and install an
|
||||||
|
MMJ loopback on both ports 2 and 3. Insert and load the scratch floppy
|
||||||
|
diskette and scratch TK50 COMPACTape cartridge from the maintenance
|
||||||
|
kit if the system has these devices installed. Run the exerciser by enter
|
||||||
|
ing TEST 101 or TEST 102. If the scratch floppy or TK50 cartridge is not
|
||||||
|
installed, the system tests those devices the same as in customer mode.
|
||||||
|
|
||||||
|
\subsubsection{System Exersiser Error Codes}
|
||||||
|
|
||||||
|
\hyperref[figure:2-8]{Figure 2-8} shows an example of the system exerciser display while running
|
||||||
|
the system exerciser in field service mode. Customer mode gives the same
|
||||||
|
display, but with a CU in place of the FS on the top of the display.
|
||||||
|
|
||||||
|
\begin{ttfig}{2-8}{System Exerciser Display Example}
|
||||||
|
KA410-A V1.0 01 FS
|
||||||
|
|
||||||
|
F 00B0 MONO 0000.0001 2 0 00:00:55.01
|
||||||
|
C 0080 DZ 0000.0001 5 0 00:01:28.03
|
||||||
|
B 0010 MEM 0175.0001 2 0 00:02:03.07
|
||||||
|
7 0090 HDC 2000.0001 4 0 00:02:08.58
|
||||||
|
? 0700.7091 0 00:02:08.58
|
||||||
|
?? 1002.0051 0 00:01:03:45
|
||||||
|
2500.0001
|
||||||
|
6 00A0 TPC 1000.0001 9 0 00:02:44.04
|
||||||
|
1300.0001
|
||||||
|
?? 1 OOCO NI 0000.7004 3 0 00:04:46.32
|
||||||
|
|
||||||
|
>>>
|
||||||
|
|
||||||
|
(The error codes are defined in the troubleshooting
|
||||||
|
procedures for each individual device)
|
||||||
|
\end{ttfig}
|
||||||
|
|
||||||
|
The first line indicates the CPU name (KA410-A), the ROM version (V1.0),
|
||||||
|
the ROM Status (01 -- the ROM is corrupted if this is anything other than
|
||||||
|
01), and the mode of the system exerciser (CU for customer and FS for field
|
||||||
|
service mode). The next several lines list information on the devices that
|
||||||
|
are exercised by the system exerciser.
|
||||||
|
|
||||||
|
The first column of the display lists the test number (F,C,B,7,6,1). TESTS 4
|
||||||
|
through 1 are option modules and these options may not be installed. They
|
||||||
|
are listed in the display only if they are installed. The second column lists
|
||||||
|
the device identifier (00B0, 0080, 0010,...). Next is the mnemonic for each
|
||||||
|
device (MONO, DZ, MEM,...) followed by the error code (0000.0000) for
|
||||||
|
that device (the HDC and TPC devices have additional error codes for each
|
||||||
|
drive). After the error code is the number of times the device was tested
|
||||||
|
followed by the time of the last pass the exerciser made on that device.
|
||||||
|
The time is in the format of days hours:minutes:seconds.hundredths of
|
||||||
|
seconds. Question marks identify hard (??) and soft (?) errors. See the
|
||||||
|
troubleshooting procedures (\hyperlink{section.2.4}{Paragraph 2.4}) for
|
||||||
|
the device that has an error detected with it.
|
||||||
|
|
||||||
|
\section{Troubleshooting Procedures}
|
||||||
|
|
||||||
|
Each section below contains troubleshooting information for the device
|
||||||
|
indicated. Find the section below with the mnemonic (such as MONO or TPC)
|
||||||
|
that you want to troubleshoot. Once the section is located, read through the
|
||||||
|
procedures to familiarize yourself with the testing of the device.
|
||||||
|
|
||||||
|
\subsection{MONO -- Monochrome Video Troubleshooting Procedures (VAXstation 2000 only)}
|
||||||
|
|
||||||
|
You can troubleshoot the monochrome circuitry on the system module
|
||||||
|
in either customer mode or field service mode. Both modes test the
|
||||||
|
monochrome circuits the same. This procedure does not troubleshoot the
|
||||||
|
video monitor, just the video circuits on the system module.
|
||||||
|
|
||||||
|
\subsubsection{Self-test}
|
||||||
|
|
||||||
|
To run self-test on MONO, enter TEST F. The monitor connected to the
|
||||||
|
video port blinks white and black several times. This is normal. The results
|
||||||
|
of the self-test are displayed when self-test is complete. Any error code
|
||||||
|
other than 0000.0001 indicates a fault in the monochrome video circuits.
|
||||||
|
You must replace the system module to fix this problem.
|
||||||
|
|
||||||
|
\subsubsection{System Exerciser}
|
||||||
|
|
||||||
|
To run the system exerciser, enter TEST 0 for customer mode or TEST 101
|
||||||
|
for field service mode. The monitor connected to the video port blinks white
|
||||||
|
and black several times while the monochrome circuits are being tested. The
|
||||||
|
results of the MONO tests are displayed on the console screen after they are
|
||||||
|
tested for the first time, then displayed again after each concurrent testing.
|
||||||
|
Any error code other than 0000.0001 indicates a fault in monochrome video
|
||||||
|
circuits. You must replace the system module to fix this problem. This
|
||||||
|
device is not exercised on the MicroVAX 2000 system since MicroVAX 2000
|
||||||
|
does not use video monitors.
|
||||||
|
|
||||||
|
\subsection{CLK -- Time-of-Year Clock Troubleshooting Procedures}
|
||||||
|
|
||||||
|
You can troubleshoot the CLK circuitry in either customer mode or field
|
||||||
|
service mode. Both modes test the CLK circuits the same.
|
||||||
|
|
||||||
|
\subsubsection{Self-test}
|
||||||
|
|
||||||
|
To run self-test, enter TEST E. Any error code other than 0000.0001
|
||||||
|
indicates a fault in the CLK circuits. An error code of 0000.0005 indicates that
|
||||||
|
the system time is not set. Refer to the operating system software
|
||||||
|
documentation to set the time. If any other error code appears in the CLK error
|
||||||
|
display, you must replace the system module.
|
||||||
|
|
||||||
|
The system exerciser does not display the Status of the CLK circuits. The
|
||||||
|
CLK circuits are not directly tested but are tested through the testing of
|
||||||
|
other circuits.
|
||||||
|
|||||||
Binary file not shown.
BIN
README.pdf
BIN
README.pdf
Binary file not shown.
Reference in New Issue
Block a user